Perkin Elmer Spectrum 400 IR/NIR#
Fourier-transform infrared spectroscopy (FTIR) is a technique used to obtain an infrared spectrum of absorption or emission of a solid, liquid or gas. An FTIR spectrometer simultaneously collects high-resolution spectral data over a wide spectral range. This confers a significant advantage over a dispersive spectrometer, which measures intensity over a narrow range of wavelengths at a time.
The term Fourier-transform infrared spectroscopy originates from the fact that a Fourier transform (a mathematical process) is required to convert the raw data into the actual spectrum.
Procedure#
- Before you start making solutions in the Pchem lab, ensure that the Perkin Elmer Spectrum 400 FT-IR/FT-NIR Spectrometer is turned on (it should not be turned off) 
- Log into the Spectrum IR software on the computer   
- The first set of samples are analyzed using the movable salt plates (in a desiccator to the left of the instrument) 
Salt Plate Analysis#
- In the Tabs on the bottom of the screen, you will be able to adjust several factors for your experiment such as: sample name, save location, instrument beam path, advanced settings, and basic instrument set-up 
- In this experiment, you will mainly be focusing on just sample name, save location, and basic instrument set-up to adjust a few parameters 
- It is also noted to adjust the parameters at the top of the page as well so that they reflect what you have included in the bottom (specifically Sample ID)   
Sample Naming#
- You can enter a name for your sample in both the Auto-name and Auto-description box after the [nnn] listed   
- Sometimes the auto-naming function does not work as you would like, so you may manually enter a name for the file up at the top of the page where it says βSample IDβ. 
Data Collection#
- Under Auto-Save Options you can chose where your files will save and the file format to export as   
Basic Instrument Settings#
- Here you can chose the wavelength range you wish to measure your samples within, the resolution, scan type, accumulations (how many scans), and units (absorbance, etc.)   
Scanning Your Samples#
- Before scanning any of your samples, you must collect a background spectrum beforehand 
- The sample compartment should contain the salt plate with your solvent. 
- Click Scan   
- Starting with the lowest concentration sample, place several drops (2 or 3) of your sample onto the salt plate using a Pasteur pipet 
- Load the salt plate into the sample holder. Be sure to only touch the outer edges of the salt plates to insert it into the holder. 
- You may notice that the βScanβ button you used has changed to a slightly different icon after running the background (it will still say Scan) 
- Click Scan and notice that a green progress bar appears at the bottom of the Viewing Pane 
- When the scan is complete, you will see your spectrum appear in the Viewing Pane 
- You can see a list of all of the scans you run below the Viewing Pane, as well as under the Results tab listed near the top   
- After each scan, remove the salt plates carefully, and clean with methylene chloride between uses. This done by adding 2 drops to each salt plate and letting the plates air dry. - Caution - The salt plates are very fragile, so when you go to clean them make sure to have a kim wipe down on the benchtop to set them on. 
- Load your next sample onto the plates and insert them into the compartment within the instrument 
- Continue this process for the rest of your samples by repeating the above steps 
Data Analysis#
- Determine the peak height (essentially the maximum absorbance) of the spectrum for each visible peak int he regions of interest 
- You can do so by either exporting your data and estimating, or you can utilize the Peak Height function within the software   
Fixed Pathlength Cell Analysis#
- Similarly, you will run your samples again using a fixed pathlength cell to determine which method is better at analyzing your samples 
- Using a plastic syringe, you will inject your sample into the cell (it may overflow slightly) 
- Following the sample procedure above, you will run each of your samples, flushing the fixed cell with methylene chloride in between runs. - Tip - Starting with the least concentrated sample and working up to the most concentrated sample will greatly increase the accuracy of the data for fixed salt plates. 
- Once all of your scans has finished you can analyze your data similar to the steps provided above when using the salt plates.